Email: xalabfurniture@163.com 丨  Phone: +86 18688931479
Multichannel Automatic Micropipette Electronic Adjustable Volume Autoclavable Lab Pipette with Self-calibration
Home » Products » Lab Accessories » Multichannel Automatic Micropipette Electronic Adjustable Volume Autoclavable Lab Pipette with Self-calibration

loading

Share to:
facebook sharing button
twitter sharing button
linkedin sharing button
line sharing button
wechat sharing button
pinterest sharing button
whatsapp sharing button
sharethis sharing button

Multichannel Automatic Micropipette Electronic Adjustable Volume Autoclavable Lab Pipette with Self-calibration

This is is an electronic pipette which is useful in almost all kinds of research institutes and laboratories. Our lab pipette is precised in operation and ergonomic in appearance.
LCD digital display with motorized operation reduces the RSI with increased productivity and offering a new pipetting experience to the users.
  • Lab pipette multichannel

  • Xunling

  • XLPM

Price:
Availability:

Multichannel Automatic Micropipette Electronic Adjustable Volume Autoclavable Lab Pipette with Self-calibration


Multi-functional high performance 8-channel electronic pipette offers productive pipetting with its easy to

understand operations. Its uniquely lightweight streamlined design ensures an effortless transfer of multiple

samples with an increased throughput and data reproducibility.

Multichannel lab pipette supplier-Xunling

Key Features

High accuracy, high performance stepper motor ensures accuracy and repeatability, eliminating manual pipetting errors

•   A motor driven digital control pipette with multifunction

•   2 buttons handle all operational settings

•   Light weight, ergonomic design, small body dimension for easy handling that guarantees fatigue free pipetting

•   Pipetting, Mixing, Stepper and Dilution

•   Adjustable speeds for aspiration and dispensing

•   Li-ion battery and dual charging modes enable longer operation time

•   Self  calibration  –  dPette  can  be  connected  to  a  PC  via.  USB  for calibration by the user. Third party calibration is not required.


Three easy steps to operate this electronic digital pipette

  1.  Long press the Parameter Knob for 2 seconds to Start.

  2. Quickly turn the Function Wheel to Activate, Rotate to Switch Pipetting, Continuous Dispensing, and other Function Settings.

  3. After quickly turning the Parameter Knob to Unlock, Turn to Adjust the Volume, Press to Pipette, Long press to Mix.

Multichannel lab pipette price-xunling

Best Lab electronic pipette

Specifications

Channels
Volume Range Increment Test Volume Accuracy Error Precision Error
μL μL μL μL % s.d.μL CV%
8 0.5-10uL 0.01
10 ±0.20 ±2.0 ±0.10 ±1.0
1 ±0.08 ±8.0 ±0.05 ±5.0
8 10-100uL 0.1
100 ±0.80 ±0.8 ±0.30 ±0.3
10
±0.30 ±3.0 ±0.20 ±2.0
8
30-300uL 1
300 ±1.80 ±0.6 ±0.90 ±0.3
30 ±0.90 ±3.0 ±0.30 ±1.0

* SD=Standard Deviation * CV=Co-efficient of Variation